UncleSteve,
You can still purchase a "new" Synrad Laser Tube if needed for approximate cost of $6000.00 to $7000.00.
This would be directly from Synrad.
From: laserman (MIKEMAC) [#9]
6 Apr 2007
To: Tony (ANTE) [#5] 6 Apr 2007
Tony,
It is possible that the tube is needing recharged, however you should adjust the tickle width and leave the KHz alone. Try this setting
5KHz and 1us.
Are you changing this on the tube or in the system menu?
From: UncleSteve [#10]
6 Apr 2007
To: laserman (MIKEMAC) [#8] 6 Apr 2007
From: David Wilhite (DAVID-ULS) [#11]
6 Apr 2007
To: ALL
Tony,
I agree with Mike, it sounds like your laser may need recharging.
Remember, you're always welcome to call the ULS technical support department at 480-609-0297 with any additional questions you may have.
All the best,
Dave Wilhite
From: Tony (ANTE) [#12]
6 Apr 2007
To: laserman (MIKEMAC) [#9] 6 Apr 2007
Mike, the changes I make are in system menu. I will try your suggestion and see what hapens.
Tony
From: Tony (ANTE) [#13]
6 Apr 2007
To: David Wilhite (DAVID-ULS) [#11] 6 Apr 2007
Thank's David,
If the tube requires rechargeing, what is turnaround timewise to have it serviced, and the cost to have it done.
Tony
From: laserman (MIKEMAC) [#14]
6 Apr 2007
To: Tony (ANTE) [#13] 6 Apr 2007
Tony,
I am attempting to assist David here because he is not a "Service Tech" he would need to ask the service dept to answer these questions.
#1 the cost to recharge the 40 watt laser tube is $950.00+shipping the lead time should not be to bad however I can not answer that question. You should contact the service dept. at 480-609-0297 they would be able to give you the shipping times. Typically if you ordered one they will send it to you and then you will send the bad one back to them and then you will be charged the re-charge cost + shipping. All of the ULS tubes are completely re-built and upgraded to current standards this includes electronics as well as the gases.
#2 change the system tickle setting to (5kHz and 0us) if the system has problems firing then you are better than 70% that the tube needs to be re-charged. If this corrects the bleeding issue then leave it at that setting.
If it does not correct the bleeding issues then we can help to adjust it on the tube to see if this corrects the problem.
EDITED: 6 Apr 2007 by MIKEMAC
From: Tony (ANTE) [#15]
6 Apr 2007
To: laserman (MIKEMAC) [#14] 6 Apr 2007
From: laserman (MIKEMAC) [#16]
6 Apr 2007
To: Tony (ANTE) [#15] 6 Apr 2007
Tony,
I completely understand the easter get together I hope you enjoy your weekend.
The latest driver version for XP is 1.07M. Let me know if you need it I can e-mail it to you.